Product overview
PCT202 Integrated Temperature Transmitter measures medium temperature by using the characteristics of platinum resistance, which changes with temperature change in a certain functional relationship. The product consists of a temperature sensor and a conversion circuit. It has advantages of stable performance, high sensitivity, and high reliability.
PCT202 adopts all-welded movable thread structure, which is easy to install. It is widely used in automatic temperature measurement and control systems for petroleum machinery, chemical machinery, pumps and compressors, electric power, boilers, and natural gas, etc.
Features
Movable connector, easy to install
High strength, high pressure resistance
Integrated structure
Long-term stable work
Fast response
Reliable structure
Applications
Equipment support
Automatic temperature measurement and control systems such as pumps, compressors, natural gas pipelines, etc.
Temperature measurement of water or oil in petroleum, chemical, electric power, textile, and environmental protection fields, etc.
Notes:
1 When measuring temperature, the temperature measuring element should reach thermal equilibrium with the measured object.
2 Make sure the insertion length is accurate.


Note: Torque depends on various factors such as material of gasket, supporting materials, lubrication of thread and pressure.


Example: PCT202-T2B1C1L050J5
Refer to product model PCT202, with temperature range -50ºC~150ºC, supply 16~36VDC, output signal 4~20mA, pressure port M20×1.5-6g, insertion length 50mm, and electrical connection DIN43650.
Ordering tips:
PCT202 Integrated Temperature Transmitter measures medium temperature by using the characteristics of platinum resistance, which changes with temperature change in a certain functional relationship. The product consists of a temperature sensor and a conversion circuit. It has advantages of stable performance, high sensitivity, and high reliability.
PCT202 adopts all-welded movable thread structure, which is easy to install. It is widely used in automatic temperature measurement and control systems for petroleum machinery, chemical machinery, pumps and compressors, electric power, boilers, and natural gas, etc.
Features
Movable connector, easy to install
High strength, high pressure resistance
Integrated structure
Long-term stable work
Fast response
Reliable structure
Applications
Equipment support
Automatic temperature measurement and control systems such as pumps, compressors, natural gas pipelines, etc.
Temperature measurement of water or oil in petroleum, chemical, electric power, textile, and environmental protection fields, etc.
Notes:
1 When measuring temperature, the temperature measuring element should reach thermal equilibrium with the measured object.
2 Make sure the insertion length is accurate.
Performance parameters | |
Temperature range | -50ºC~250ºC |
Supply & output | 4~20mA(16~36VDC) |
Accuracy | ±0.5%FS(typ.) |
Sensing element | Pt100 Grade A |
Insulation strength | 100MΩ/250VDC |
Response time | ≤30s |
Long term stability | ≤0.2%FS/year |
Insertion diameter | Φ8mm |
Storage temp. | -40ºC~85ºC |
Electrical connection | DIN43650, DIN43650 with cable |
Protection | IP65 |
Material | 304 |
Max. mounting torque | 25Nm |
Structure |

Pressure port |

Note: Torque depends on various factors such as material of gasket, supporting materials, lubrication of thread and pressure.
Temperature range selection | ||
Temperature range code | Temperature range | Remark |
T1 | -50ºC~100ºC | |
T2 | -50ºC~150ºC | |
T3 | -50ºC~200ºC | |
T4 | -50ºC~250ºC | |
T5 | 0ºC~100ºC | |
T6 | 0ºC~150ºC | |
T7 | 0ºC~200ºC | |
T8 | 0ºC~250ºC |
Accessory |

How to order |

Example: PCT202-T2B1C1L050J5
Refer to product model PCT202, with temperature range -50ºC~150ºC, supply 16~36VDC, output signal 4~20mA, pressure port M20×1.5-6g, insertion length 50mm, and electrical connection DIN43650.
Ordering tips:
- Ensure compatibility between measured media and contacting part of product.
- For special requirements on the product appearance and performance parameters, customization is available.